Impedance of Protective Earth Connections
The Resistance was measured according to IEC 60601-1 3rd Edition but with 40A test current instead of 25A, with the limit 0.1 Q for mains connection with removable power cords. Resistance was calculated using the formula R = U/I (Measured resistance is “Measured Position” to ground earth pin of mains inlet via a mains connector)
Proposed translations (Russian)
3 ->
Change log

Jan 3, 2021 08:46: changed "Kudoz queue" from "In queue" to "Public"

Proposed translations

31 mins
Selected

->

to ground earth pin of mains inlet via a mains connector =
соединение на массу клеммы заземления входа питания от сети через сетевую розетку

Т.е., заземление устройства производится не через отдельный провод заземления, а через заземляющий провод и сетевую розетку

--------------------------------------------------
Note added at 37 mins (2021-01-03 09:23:51 GMT)
--------------------------------------------------

Подключение заземления
https://kzref.org/rukovodstvo-po-installyacii-i-ustanovke-so...
Заземление через сетевую розетку: Для этой цели можно использовать только розетки с заземляющим контактом. Для проверки надежности заземления пользуйтесь трехпроводным тестером...

--------------------------------------------------
Note added at 41 mins (2021-01-03 09:27:25 GMT)
--------------------------------------------------

Заземление устройства используется не отдельная шина заземления, а заземляющий провод шнура питания через контакт заземления в сетевой розетке